Ultra-low Refrigerated Circulator with 2-stage cascade-cooling
JULABO Ultra-Low Refrigerated Circulators for heating and cooling are suitable for external temperature tasks. Powerful circulating pump systems and high heating and cooling capacities guarantee short heat-up and cool-down times. ACC ‘Active Cooling Control‘ provides active cooling control across the whole temperature range. An energy saving feature with only minor loss of heat is provided by the proportional cooling capacity control in the ‘FP’ refrigeration units. The inevitable dust accumulation that occurs in refrigeration systems is solved by means of an easily removed venting grill, which allows for easy cleaning. The units are equipped with handles or castors for an easy transportation. All models have a drain tap on the front allowing the fluid to be drained easily. Additionally the instruments are equipped with an improved insulation to help avoid ice-formation, and have a visual liquid level display.
| Order No. | 9162689 |
|---|---|
| Model series | TopTech |
| Category | Ultra-Low Refrigerated-Heating Circulators |
| Working temperature range (°C) | -90 ... 100 |
| Temperature stability (°C) | ±0.02 |
| Setting / display resolution | 0.01 °C |
| Integrated programmer | 1x10 steps |
| Temperature Display | VFD |
| Heating capacity (kW) | 1.3 |
| Cooling capacity (Medium Ethanol) | |
| 20°C | 1 kW |
| 0°C | 0.92 kW |
| -20°C | 0.88 kW |
| -40°C | 0.75 kW |
| -60°C | 0.58 kW |
| -80°C | 0.2 kW |
| Pump capacity flow rate (l/min) | 11-16 |
| Pump capacity flow pressure (bar) | 0.23-0.45 |
| Bath opening / bath depth (W x L / D cm) | 13 x 15 / 16 |
| Pump connections | M10x1 |
| Barbed fittings diameter (inner dia. / mm) | 8 / 10 |
| Filling volume (liters) | 8 |
| Refrigerant | R404A R508B |
| External Pt100 sensor connection | integrated |
| Digital interfaces |
RS232 Optional Profibus |
| Ambient temperature | 5...40 °C |
| Dimensions W x L x H (cm) | 55 x 60 x 90 |
| Weight (kg) | 133 |
| Classification according to DIN12876-1 | Classification III (FL) according to DIN 12876-1 |
| Included with each unit | 2 each barbed fittings for tubing 8 and 10 mm inner dia. (pump connections M10x1 female). |
| Cooling of compressor |
Air |
Condensation trap to prevent ice formation available as accessory.
Technical changes without prior notification.
Images may deviate from the original.
We would be pleased to submit a non-binding offer for the selected product or answer all pertinent questions.